Saved settings – Ocean Optics NeoFox User Manual
Page 43
3: Using NeoFox with NeoFox Viewer Software
013-20000-009-02-201203
33
2. Wait long enough for the temperature and tau values to settle – this amount of time will vary
based on the probe type and the number of averages parameter. You should wait at least one
3.
Single Point Reset button for at least five seconds. The Single Point Reset control button
is recessed in the center back of the NeoFox unit.
the temperature probe or the
ave selected for your device.
eans that all
minute.
Push the
After five seconds, NeoFox will use the following three values to perform the single point reset:
Tau will be the current value of Tau measured from the probe.
Temperature will be either the current temperature reading from
fixed temperature depending on which temperature source you h
If no temperature is selected, the SPR will not occur.
Percent oxygen will be the value that was set during the setup phase in the lab.
Once the Single Point Reset has been performed, NeoFox saves your settings. This m
of the parameters listed in
Saved Settings
will be saved in their current states.
Saved Settings
d onboard the NeoFox in its non-volatile memory. When the settings
are saved, the saved values will be loaded when the device turns on:
n
tion Factor
ttings
e
alue
eset Settings
en
ture
The following parameters are save
Measurement Parameters:
Duty Cycle Seconds o
Duty Cycle Seconds off
Number of Averages
Oxygen Units
Salinity Correc
Calibration Settings:
Temperature Se
Temperature Sourc
Fixed Temperature V
Calibration Method
Sensor Type
Single Point R
Single Point – Tau
Single Point – Oxyg
Single Point – Tempera