beautypg.com

Viewing the data log file name, Setting the data logging interval, Data log file format – Ocean Optics NeoFox User Manual

Page 56

background image

4: Using NeoFox Sport

46

013-20000-009-02-201203

Viewing the Data Log File Name

You cannot set the file name, but you can view the name of the current log file by selecting Oxygen
Sensor | Sensor Setup | Data Logging | Data log filename.

Setting the Data Logging Interval

You can also set the data logging interval via the Oxygen Sensor | Sensor Setup | Data Logging | Data log
interval menu selection. This can be used to reduce the size of the log file. The minimum data log interval
is 1 second.

Data Log File Format

The datalog file is formatted as a tab separated value file. This file can be easily imported into Microsoft
Excel or comparable spreadsheet programs. A single sample in the data log is formatted as follows.

millisecond count upload date upload time oxygen
oxygen unit label algorithm temperature sensor
temperature tau algorithm pressure sensor pressure

Where:
Millisecond Count = The millisecond counter of the NeoFox device at the time of the actual sample.
Because this value is stored at the actual sampling time, this value is a more reliable timing metric than
the “upload date” and “upload time” fields.

Upload Date / Upload Time = These fields represent the date and time that the sample was uploaded to
the DPU.

Oxygen = The oxygen measurement (in whatever units are currently selected).

Oxygen Unit Label = The oxygen unit label; either “ppm”, “%”, “Torr”, or “um/L”

Algorithm Temperature = The temperature that was used to perform the conversion from Tau to Oxygen.

If the NeoFox’s temperature source parameter is set to:

Sensor temp, then this is the sensor temperature.

Fixed temp, then this is the fixed temperature value.

None, then this is 200.

Sensor Temperature = The sensor temperature (in degrees C).

Tau = The tau value (in microseconds).

Algorithm Pressure = The temperature that was used to perform the conversion from partial pressure of
Oxygen to concentration or mole%.

If the NeoFox’s pressure source parameter is set to:

Sensor Pressure, then this is the sensor pressure.

Fixed Pressure, then this is the fixed pressure value.

None, then this is 101 – 325.

Sensor Pressure = The pressure (in kPa).