beautypg.com

Ocean Optics NeoFox User Manual

Page 41

background image

3: Using NeoFox with NeoFox Viewer Software

013-20000-009-02-201203

31

Procedure

To perform a Single Point Reset,

1. From the Main screen, select Options | Calibration.

2. Click

the

Setup tab.

3. Under Select Sensor Calibration Method, choose Single Point Calibration.

4. Under Select How Temperature Will Be Measured, choose Use NeoFox Temperature Sensor.

5. Select

the

Single Point Calibration tab. The Single Point Calibration screen appears.

6. Single point calibration is always performed within the calibration range. This is usually done in

ambient conditions at room temperature. Enter the known oxygen concentration the sensor is
exposed to in the Oxygen field. Enter the known temperature reading from the oxygen sensor in
the Temperature field. (The NeoFox temperature reading can be observed on the main NeoFox
Viewer window.)

7. Once the sensor is in stable condition with a stable Tau value, click on Use Current Tau.

8. To use this calibration for oxygen measurement, click on the graph, then click the

button. A Download Confirmation screen appears.

9. Click

Yes to confirm the download and continue.