Ocean Optics NeoFox User Manual

Page 36

background image

3: Using NeoFox with NeoFox Viewer Software

26

013-20000-009-02-201203

Procedure

1. From the Main screen, select Options | Calibration.

2. Click

the

Setup tab. The Setup screen appears.

3. Under Select Sensor Calibration Method, choose Multi-Point Calibration.

4. Under Select How Temperature Is Measured, choose Use NeoFox Temperature Sensor.

5. Under Enter Sensor Type And Optional Data, choose the type of sensor and add the serial number

and any notes needed.

Note

Normalized decay time = τ0/τ.

6. Click

the

button, then browse to the location of the factory calibration file on the

NeoFox CD.

7. Select

the

Multi-Point Calibration tab. The uploaded calibration table appears on the left side of

the screen. The curve-fitted graphs appear on the right side of the screen.