Page 38

background image

3: Using NeoFox with NeoFox Viewer Software

28

013-20000-009-02-201203

3. Under Select Sensor Calibration Method, choose Two-Point Calibration.

4. Under Select How Temperature Will Be Measured, choose None (Two point calibration only).

5. Under

Enter Sensor Type And Optional Data, choose the type of sensor and add the serial

number and any notes needed.

6. Select

the

Two-Point Calibration tab. You need two known concentrations of oxygen to perform

this calibration, one being the 0% oxygen concentration.