5 error codes – VEGA VEGAMET 514… D User Manual
Page 49
VEGAMET 514… D
49
5.5 Error codes
Error code
Meaning
E004
This operation is not supported by the instrument (different software version). You tried to
activate a function with a newer software version of VEGA Visual Operating in an instrument
with older software version. However the instrument cannot support this operation.
Please update the software of your instrument.
E007
Sensor type not suitable.
A VBUS-sensor with the serial number stated by you was found, however the sensor does
not correspond.
Please check the adjusted serial number or change the configuration.
E008
VBUS-sensor not found. The sensor cannot be found.
Reasons:
- sensor is not connected
- no sensor connected with this serial number
- line break (VBUS)
- shortcircuit (VBUS)
- the sensor is defect
- interfered communication (VBUS)
probably
- interference in the VBUS-transmitter/receiver part of VEGAMET
Please check the serial number, the connection line, the sensor and if necessary the
VEGAMET.
E010
The serial numbers of the connected VBUS-sensor has not been coordinated.
E013
VBUS-sensor signals failure. The VBUS-sensor signals failure; the communication with the
sensor however is okay.
Reasons:
- failure on the VBUS-sensor
Please check the VBUS-sensor.
E014
Suspected shortcircuit
Reasons:
- shortcircuit
- sensor is defect
Please check the connection line and if necessary the sensor.
E015
Suspected line break
Reasons:
- no sensor connected
- line break
- sensor is defect
Please check the connection line and if necessary the sensor.
E016
Empty/full adjustment exchanged. The values for the empty adjustment are higher than the
values for the full adjustment. Please carry out adjustment again.
E017
Adjustment span too small. The adjustment values are too close togther or are identical.
Please carry out adjustment again.
E021
Span of the sensor characterstic values too small or values exchanged. The sensor
characteristic values are too close together or are identical or the min. sensor characteristic
value is higher than the max. sensor characteristic value. Please check the sensor
charactersistic values.
Diagnosis