VEGA VEGAMET 514… D User Manual
Page 36
36
VEGAMET 514… D
Set-up, parameter adjustment
Parameter adjustment MST1 and MST2, outputs
Current output 3 (continuation)
–– Pulse parameter
–– Volume
1000
Range 0 … 9999
–– Quantity x factor
1
Multiplication factor x1, x10, x100, x1000
–– Creeping quantity
suppression
0 %
Range 0 % … 100 %
–– Sampling
h 1
Range 1 … 100 h (hours)
This menu point is only displayed in mode "Sampling
output".
–– Volt outputs
–– Volt output 1
acc. to basic configuration coordinated to MST1, i.e. 0 … 10 V relating
to level
–– relating to
Percent
etc. choice acc. to list
–– Current output
2/10 V
0/10 V
10/2 V
10/0 V
-free-
activates the menu points "Current at 0% and 100%"
–– Volt at 0 %
2 V
adjustable range 0.0 … 10.0 V
–– Volt at 100 %
10 V
adjustable range 0.0 … 10.0 V
Reaction of the volt output in case of failure
–– at failure
0 V
0 V
11 V
11 V
--
Actual value is kept
0 %
Volt value relating to 0 %
100 %
Volt value relating to 100 %
–– Volt output 2
as volt output 1, however acc. to basic configuration
coordinated to MST2, i.e. 0 … 10 V relating to flow.
Volt output 2 can be configured to level like volt output 1.
–– Volt output 3
as basic configuration coordinated to no MST (– –).