Other Field Adjustments
Once EC coulter depth and sampling depth are satisfactory, adjust other components in this
sequence:
1. Scraper adjustment: in manual mode hydraulically raise the sampling shoe to maximum
height. Adjust scraper until cutting shoe clears scraper blade when sampler shank is fully
raised.
2. Adjust electrode holder: with sampling mechanism raised completely, adjust electrode holder
to provide ½” (1.2 cm) clearance between it and sampling trough.
3. Wash adjustment: Wash brackets should be parallel to sampling trough, with jets directly
beside electrode holder, jets should be 1” (2.5 cm) below electrodes; when electrodes are
properly aligned.
Turn crank to raise or
lower electrode holder
electrode
holder 1”
(2.5cm)
above wash
jets
Adjust scraper bracket until
cutting shoe clears scraper
when sampler assembly is
raised completely.
½” (1.2 cm) clearance
between electrode
holder and trough liner