beautypg.com

Veris Technologies MSP EC and pH (1997-2008 CF card drive 1.10M) - Operating Instructions User Manual

Page 24

background image

Veris Technologies

Pub. #OM MSP03-1

24


Connecting cables to External controller as shown below:

Remove pH electrodes from individual storage containers and fill soaker solution container with
soaker solution. Install soaker solution container on electrode holder. Loosen plastic set screws on
electrode holder and insert pH electrodes into electrode holder. Re-tighten set screws finger tight and
lock in place with lock nuts. Do not overtighten set screws or electrode damage may occur. Always
keep electrodes in soaker solution, either in individual containers or soaking in large container
installed over electrode holder. Route electrode cables away from sampling mechanisms to prevent
damage

—tie-strap excess length of cable as needed.

ball valve:
closed
open

pH
electrodes

electrode set
screws and
lock nuts

electrode
holder

soaker
container