CKW-55 CHECKWEIGHER

- - - - - SPAN CAL - - - - -
Cal: 40 kg
Old: 39.99 kg
Diff: 0.01 kg
Wt. Ref. _____________

12/01/04 1:10:00 PM
USER ID 654321
PROJ ID 111111
SCALE ID 123456
Name _______________
- - - - - END - - - - -

- - - - - CAL TEST - - - - -
Test: 10 kg
Act: 10.01 kg
Diff: 0.01 kg
Wt. Ref. _____________

12/01/04 1:30:00 PM
USER ID 654321
PROJ ID 111111
SCALE ID 123456
Name _______________
- - - - - END - - - - -

- - - - - LIN CAL - - - - -
Cal: 40 kg
Old: 39.99 kg
Diff: 0.01 kg
Wt. Ref. _____________

12/01/04 1:10:00 PM
USER ID 654321
PROJ ID 111111
SCALE ID 123456
Name _______________
- - - - - END - - - - -

4.1.5 GMP Calibration Printout

If any setting in the Print-Content-GMP menu (refer to Section 3.6.4 Print Menu) is turned on, GMP automatically prints the
calibration data after calibration or calibration test is completed. Other GMP data, if turned on, will also be printed as shown.

Span Calibration Printout

When performing a Span calibration with any GMP setting turned on, a printout is
automatically made after the procedure is completed.

Linearity Calibration Printout

When performing a Linearity calibration with any GMP setting turned on, a printout is
automatically made after the procedure is completed.

Calibration Test Printout

When performing a Calibration Test with any GMP setting turned on, a printout is
automatically made after the procedure is completed.