beautypg.com

Ohaus CKW WASHDOWN CHECKWEIGHING SCALES_INDICATOR Manual en User Manual

Page 18

background image

EN-14

CKW-55 CHECKWEIGHER

Auto-Tare

Auto-Tare automatically tares the initial weight (such as a container) placed on the empty platform, without having to press
the TARE button. The tare value is cleared automatically when the weight on the pan is fully removed.

In Auto-Tare mode, the Tare indicator blinks when waiting for a Tare item to be placed on the empty scale platform (with
LFT On, there is no flashing Tare indicator).

When the On Accept setting is selected, weight values that are within the accept range will be tared automatically.
Note: Auto-Tare supersedes any pre-set (PT) value in memory.

Displaying Gross, Net, Tare

When a tare has been entered, pressing the G/N/T

Menu button successively displays the Tare, the Gross, and the Net

weights. After a few seconds the scale automatically reverts to displaying the NET weight.

Changing Units of Measure

Press and hold the PRINT

Units button until the desired measuring unit appears. Only measuring units enabled in the

Readout-Unit Menu will be displayed (refer to Section 3.6.2).

Printing Data

Printing the displayed data to a printer or sending the data to a computer requires that the communication parameters in the
Print and Communication Menu are set (refer to Sections 3.6.4 and 3.6.5).

Press the PRINT

Units button to send the displayed data to the RS232 port (the Auto-Print Mode in Section 3.6.4 function

must be Off).

Note: If the PRINT

Units button is pressed too long, the display advances to another measuring unit without sending the

print data.