M&C TechGroup KS 2.Ex Data sheet User Manual
Liquid alarm sensor ks 2 / ks 2.ex, For non-conductive and conductive media, Special features
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
8.3
Embracing Challenge
Sure detection even of minimal liquid
quantities
Integrable in M&C-universal filters and
flow chamber
Explosion proof version according to
ATEX for zone 0
Special Features
Liquid alarm sensor KS 2 / KS 2.Ex
for non-conductive and conductive media
KS 2
Application
Liquid alarm sensors are used in gas sam-
ple conditioning systems for monitoring gas
cooling and condensate drain-off devices in
order to provide protection for downstream
analysis instruments. This monitoring device
KS 2/KS 2.Ex reliably signals the penetration
of non conductive (e.g. alcohol) and conduc-
tive (e.g. water) liquid in the event of cooling
or condensate drain-off equipment being
defective, thus avoiding expensive down time
as well as high repair costs for analysis instru-
ments.
In the event of an alarm, power for the sample
gas pump must be switched off or a sole-
noid valve must be switched idle in the sam-
ple conditioning system.
Description
The M&C liquid sensor KS 2/KS 2.Ex works on
the principle of capacitive measurement and
is suitable for non-conductive and conduc-
tive media. A pre-amplifier is integrated in the
sensor housing and is connected with the
necessary external electronic controller via
2- resp. 3-wire. For KS 2 the required electron-
ic controller is available in various versions,
FA1.1; FA1.4 and is described in the separate
data sheet 9.5.
The M&C liquid sensor KS 2.Ex for use in
hazardous areas and media should only be
applied in connection with electronic control-
ler WE77.Ex1 or KFA6-SR2-Ex1W.
The M&C KS 2/KS 2.Ex liquid sensor is con-
structed in such a way that any droplets of
liquid in the sample gas are attracted directly
to the active sensor surface. Even the smallest
liquid droplets will trigger a sure and rapid
alarm.
The sensors are mounted with the 16mm ø
stainless steel body in the GL-25 connector
of the universal filter F..-..-D or the condensate
pot TG1 or in the flow chamber LS/KS.
Technical specifications and illustrations are without
obligation, subject to modifications. 04.99/11.06