M&C TechGroup SP2500-H_C_I_BB_F Data sheet User Manual

Gas sample probe series sp, Special features, Application

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

2.4

Embracing Challenge

ƒ

ƒ

Extractable sample tube or prefilter,
without dismounting the probe

ƒ

ƒ

Integrated back purge possibility with
closing off the sample gas outlet

ƒ

ƒ

Optimum operational reliability

ƒ

ƒ

Universal applicability

ƒ

ƒ

Adaption to nearly all process condi-
tions due to its compact and modular
design

ƒ

ƒ

Easy mounting

ƒ

ƒ

Problem-free maintenance

ƒ

ƒ

Small stagnant space

Special Features

Gas Sample Probe Series SP®

Version SP2500-H, SP2500-H/C/I/BB, SP2500-H/C/I/BB/F

SP2500-H/C/I/BB/F/1K190

Application

The gas sample probes M&C type SP2500...,

based on version SP2000-H (see data sheet

2.2.2), are used for continuous gas sampling

in processes with gases of high dust con-

tent, high temperatures and/or high humid-

ity. They make it possible to remove e.g. for

cleaning purposes the preliminary filter or

the sample tube out of the process without

dismounting the complete gas sample probe.
The probes type SP2500-H/C/I/BB and

SP2500-H/C/I/BB/F with special back purge

possibility are used in case of very high dust

content. For this purpose, they are equipped

with an additional back purge valve and a

pneumatic check valve in the sample gas

outlet.

Description

The M&C gas sample probes are designed for

easy installation, reliable operation, trouble-

free maintenance and universal applicabil-

ity. Depending on the application, different

sample tubes or prefilters (see data sheets

2-1.1.0.6 and 2-1.1.0.8), not included in the

scope of delivery of the probe, are screwed

into the thread (G3/4”i) of the filter housing.
The large surface ceramic filter element (also

glass fiber elements or spun glass fillings

are available) is placed in a housing with low

stagnant space outside the process.
The M&C gas sample probes are designed

so that changing a filter element does not

involve the use of tools. In this operation the

sample line does not need to be removed,

thus avoiding contamination of the clean gas

path and also maintaining the integrity of the

system.
The cleaning of the sample tube or the pre-

liminary filter can be effected by extracting

the filter from the probe.

The special design of the heating element of

the SP2500... (with protective hood) permits

controlled heating of the complete filter

housing, including the mounting flange up

tp 180 °C. This ensures reliable operation

external to the process preventing the tem-

perature falling
below the dew-point.
The temperature of the standard probe is

controlled by an integral compact design

capillary sensor thermostat with an excess

temperature limit switch and an alarm

function for temperature failure. The probe

SP2500-H provides the possibility to feed the

test gas /C optionally via a check valve.
Additional functions of the probe SP2500-

H/C/I/BB(/F):

Calibration gas will be injected into

the probe through a check valve /C

directly to the sample outlet. No cali-

bration gas is lost into the stack.

An isolation valve with pneumatic con-

trol /I shuts off the sample outlet from the

internal filter area.

Through a high flow rate check valve /

BB, which is fixed to the internal probe

area, blow back of the filter housing

area and the insitu probe tube will be

done.

Through a high flow rate check valve

/BB/F, which is fixed to the filter hous-

ing wall, blow back of the incorporat-

ed ceramic filter will be done incl. the

filter area and the insitu probe tube.

Technical specifications and illustrations are without

obligation, subject to modifications. 03.03/06.06