Technical data, Dimensions – M&C TechGroup SP2100-H Data sheet User Manual
Page 2
![background image](/manuals/256023/2/background.png)
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.8
Technical Data
Technical specifications and illustrations are without
obligation, subject to modifications. 10.04/06.06
Probe series SP
®
compact version SP 2100-H
Part No.
10S1000
10S1100
Sample tube
Type SS, stainless steel 316Ti, length 1 m*
no
Sample temperature
max. 600 °C*
Sample pressure
0.4 to 2 bar abs.
Ambient temperature
-20 °C to +60 °C
Recommended for dust loadings
up to 2 g/m
3
*
Filter chamber volume
120 ml
Filter element
S-2K 150, filter porosity 2 µm, ceramic
Probe heating temperature
+180 °C self-regulating
Ready for operation
after 2 hours
Temperature alarm contact, alarm point
<160 °C, NO
Temperature alarm, contact rating
250V-3A AC, 30V-3A DC
Connection sample outlet
1/4"-NPT inside, for max. 10 mm tube connectors
Power supply
110 up to 240V 50 / 60Hz
Power consumption
start up: 400VA, usual: 100VA, (fuse 6A)
Electrical connection
terminals max. 2,5 mm
2
, 2x PG11 cable gland
Electrical equipment standard
EN 61010, EN 60335-1
Degree of protection
IP54, EN 60529
Mounting flange
DN65 PN6,B stainless steel 316Ti
Material of sample contacting parts
SS 316 / 316Ti, FPM, ceramic
Weight
9 kg
Options
10S9005
Calibration flange, DN65PN6 with 1/8“NPT connection inclusive flange gasket and screw set
M12x80.
*
Standard, other versions on request.
∆P and T90 at flow of:
100
200
500
1000
Nl/hr
∆P pressure loss with new filter element S-2K150
7
11
20
35
mbar
T
90
time – without sample tube/prefilter –
6,0
3,5
1,0
< 0,5
sec
Dimensions
140
ш 160
1000
280
ш 25
sample out
sample out
Dimensions in mm