M&C TechGroup SP2100-H Data sheet User Manual
Gas sample probe series sp, Electrically heated, compact version sp2100-h, Special features
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.8
Embracing Challenge
Sampling of dust loaded process gases
Small volume, fast response time
Self-regulated electrical heater
Alarm contact for low temperature
Insitu probe tube 1 m long
Easy mounting and maintenance
Special Features
Gas Sample Probe Series SP®
Electrically heated, compact version SP2100-H
SP2100-H
Application
The patented M&C gas sample probe ver-
sion SP2100-H is used for continuous
gas sampling.
The compact design requires only limited
space. It is to be mounted in a weather-proof
position.
Description
The design of the M&C probe version
SP2100-H guarantees simple mounting, safe
operation and problem-free maintenance.
Changing the external filter element needs
no tools and no disassembling of the sample
line. The complete filter assembly is removed
out of the probe head. This simplifies the
checkup of the filter element and the gaskets.
The filter area of the probe head respec-
tively the insitu probe tube can be cleaned
very easely without dismounting the sample
probe.
The large external filter element of ceramic
with 2 µm porosity is heated up with the
stainless steel probe head to constant 180°C,
using a special self-regulated heating ele-
ment. This avoids dew point problems at the
probe flange and filter area.
A heat insulation covers all sides of the probe
and ensures an optimized heat distribution as
well as safe operation.
The power supply of the probe can be in
a range between 110V and 240V without
switching.
No thermostat or temperature limiter is nec-
essary. The separate thermo switch enables
a low temperature monitoring (<160°C, NO).
For electrical connection a junction box with
terminals is mounted.
A one-meter stainless steel sampling tube is
provided. This is screwed into the mounting
flange. The maximum operating temperature
of the tube out of stainless steel is 600°C.
In case to avoid cooling and condensation of
the sample in the insitu tube from the sample
point to the heated sample probe the SP
30-H or SP 35-H heated sampling tubes can
be used.
The wide range of M&C probe accessories
offers various insitu probe tubes and pre-fil-
ters to solve your specific sampling problem.
Technical specifications and illustrations are without
obligation, subject to modifications. 10.04/06.06