Technical data – M&C TechGroup PMA 10 Data sheet User Manual
Page 3

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
13.1
Technical specifications and illustrations are without
obligation, subject to modifications. 05.97/08.09
Technical Data
Version PMA 10 portable oxygen analyser
Part No.
01A1000 : PMA 10, power supply 230V 50Hz, output signal 0-1 V; 01A1000a = 115V 60Hz
Measuring ranges
selectable for 0-3, 0-10, 0-30 and 0-100 vol.% O
2
, linear
Indication
analogue / digital meter:
analogue meter selectable for each range with a scale of 0-30 and 0-100%
digital meter, 3
1
/
2
digit 9 mm high LCD for 0-100 % O
2
reading, selectivity 0,1 vol.% O
2
Output signal
0-1V DC, non-isolated, load > 100 KΩ, for each selected range;
option: 0-20mA* or 4-20mA* for each selected range, non-isolated, max. load 300Ω Part no. 01A9000
Response time for 90% FSD
< 3 seconds at 60 Nl/hr air
Accuracy after calibration
analogue = ± 1% of span / digital = ± 0,1 vol.% O
2
error of precision
Reproducibility
analogue = < 1% of span / digital = ± 0,1 vol.% O
2
error
Influence of ambient temperature
zero point ± 0,02 vol.% O
2
/ °C; sensitivity ± 0,1 vol.% O
2
/ °C
Influence of barometric pressure
The oxygen reading varies in direct proportion to changes of the barometric pressure.
Influence of sample gas flow
variation in gas flow between 0-60 Nl/hr air will cause a difference of < 0,1 vol.% O
2
.
Sample gas inlet pressure
0,01 up to 1 bar g, (PMA 10 required admission pressure for competent flow rate, no pump inside)
option: PMA 10 with internal pump, capacity 0,9 Nl/min. without pressure, Part no. 01A9100
Sample gas outlet pressure
Outlet of analyser must discharge freely into atmosphere.
Flow rate of sample gas
max. 60 Nl/hr air, adjustable with needle valve on the flowmeter 7-70 Nl/hr
Temperature of sample gas
-10 °C up to +40 °C, dry gas
Analyser temperature
according to ambient temperature, non heated version
Ambient temperature
-10 °C up to +55 °C
Storage temperature
-20 °C up to +60 °C, relative humidity 0-90% RH
Power supply
internal power unit for 230V
AC
standard or 115V
AC
available (a)* +/-10%, 40-60Hz, 8VA
option: rechargeable battery, recharged by the internal power unit, Part no. 01A9050
battery capacity 10 hr, recharge time of battery 14 hr with instrument off
Electrical connections
mains supply: 3-pole chassis plug with 2m of cable; signal: 3-pole plug
Materials in contact with sample gas
Platinum, Glass, Polypropylene, Stainless Steel 316, FPM, Epoxy resin
Sample gas connection
PP-hose nipple for DN 11-4 mm tube
Protection / electrical standard
IP 41 EN 60529 / EN 61010
Housing / colour
portable plastic housing / gray
Dimension / weight
height 150 mm, width 202 mm, depth 260 mm / approx. 3 kg
Alarms
option: O
2
alarm low* or high* adjustable 0-100% O
2
, relays contact NC/NO Part no. 01A9150
additional with acoustic signal and auto reset after 30 sec. Part no. 01A9155
additional with acoustic signal and manual reset, Part no. 01A9156
Certificate
option: approved according to DIN EN 14181 resp. to 13th and 17th BImSchV and TA-Luft,
Part no. 01A9160
* Please specify with order.
In case the options alarm and pump are both required, a rechargeable battery is automatically obliged.
In case the option “acoustic alarm” required, the option “O
2
alarm” is obliged.
Option “TÜV approved” includes 0*/4*-20mA signal output, as further options only rechargeable battery and internal pump available.
W A R N I N G ! I M P O R T A N T !
An external fine filter must always be used at the gas inlet of the analyser.
Depending on the composition of the sample gas, it may be necessary to use a
sample conditioning system. Without precautions, the analyser is only suitable for
measuring of non-hazardous gases or gasmixtures in non-hazardous areas.