Technical data – M&C TechGroup PMA 10 Data sheet User Manual

Page 3

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

13.1

Technical specifications and illustrations are without

obligation, subject to modifications. 05.97/08.09

Technical Data

Version PMA 10 portable oxygen analyser

Part No.

01A1000 : PMA 10, power supply 230V 50Hz, output signal 0-1 V; 01A1000a = 115V 60Hz

Measuring ranges

selectable for 0-3, 0-10, 0-30 and 0-100 vol.% O

2

, linear

Indication

analogue / digital meter:
analogue meter selectable for each range with a scale of 0-30 and 0-100%
digital meter, 3

1

/

2

digit 9 mm high LCD for 0-100 % O

2

reading, selectivity 0,1 vol.% O

2

Output signal

0-1V DC, non-isolated, load > 100 KΩ, for each selected range;
option: 0-20mA* or 4-20mA* for each selected range, non-isolated, max. load 300Ω Part no. 01A9000

Response time for 90% FSD

< 3 seconds at 60 Nl/hr air

Accuracy after calibration

analogue = ± 1% of span / digital = ± 0,1 vol.% O

2

error of precision

Reproducibility

analogue = < 1% of span / digital = ± 0,1 vol.% O

2

error

Influence of ambient temperature

zero point ± 0,02 vol.% O

2

/ °C; sensitivity ± 0,1 vol.% O

2

/ °C

Influence of barometric pressure

The oxygen reading varies in direct proportion to changes of the barometric pressure.

Influence of sample gas flow

variation in gas flow between 0-60 Nl/hr air will cause a difference of < 0,1 vol.% O

2

.

Sample gas inlet pressure

0,01 up to 1 bar g, (PMA 10 required admission pressure for competent flow rate, no pump inside)
option: PMA 10 with internal pump, capacity 0,9 Nl/min. without pressure, Part no. 01A9100

Sample gas outlet pressure

Outlet of analyser must discharge freely into atmosphere.

Flow rate of sample gas

max. 60 Nl/hr air, adjustable with needle valve on the flowmeter 7-70 Nl/hr

Temperature of sample gas

-10 °C up to +40 °C, dry gas

Analyser temperature

according to ambient temperature, non heated version

Ambient temperature

-10 °C up to +55 °C

Storage temperature

-20 °C up to +60 °C, relative humidity 0-90% RH

Power supply

internal power unit for 230V

AC

standard or 115V

AC

available (a)* +/-10%, 40-60Hz, 8VA

option: rechargeable battery, recharged by the internal power unit, Part no. 01A9050
battery capacity 10 hr, recharge time of battery 14 hr with instrument off

Electrical connections

mains supply: 3-pole chassis plug with 2m of cable; signal: 3-pole plug

Materials in contact with sample gas

Platinum, Glass, Polypropylene, Stainless Steel 316, FPM, Epoxy resin

Sample gas connection

PP-hose nipple for DN 11-4 mm tube

Protection / electrical standard

IP 41 EN 60529 / EN 61010

Housing / colour

portable plastic housing / gray

Dimension / weight

height 150 mm, width 202 mm, depth 260 mm / approx. 3 kg

Alarms

option: O

2

alarm low* or high* adjustable 0-100% O

2

, relays contact NC/NO Part no. 01A9150

additional with acoustic signal and auto reset after 30 sec. Part no. 01A9155
additional with acoustic signal and manual reset, Part no. 01A9156

Certificate

option: approved according to DIN EN 14181 resp. to 13th and 17th BImSchV and TA-Luft,
Part no. 01A9160

* Please specify with order.

In case the options alarm and pump are both required, a rechargeable battery is automatically obliged.

In case the option “acoustic alarm” required, the option “O

2

alarm” is obliged.

Option “TÜV approved” includes 0*/4*-20mA signal output, as further options only rechargeable battery and internal pump available.

W A R N I N G ! I M P O R T A N T !

An external fine filter must always be used at the gas inlet of the analyser.

Depending on the composition of the sample gas, it may be necessary to use a

sample conditioning system. Without precautions, the analyser is only suitable for

measuring of non-hazardous gases or gasmixtures in non-hazardous areas.