beautypg.com

M&C TechGroup PMA 10 Data sheet User Manual

Oxygen analyser series pma, Special features, Application

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

13.1

Embracing Challenge

ƒ

ƒ

Portable and rugged instrument, light-
weight

ƒ

ƒ

Analogue and digital linear indication

ƒ

ƒ

Physical measuring principle

ƒ

ƒ

Small stagnant volume, fast response
time

ƒ

ƒ

High accuracy and reliability

ƒ

ƒ

Approved according to DIN EN 14181
resp. to 13th and 17th BImSchV and TÜV
certificate TA-Luft as option

Special Features

Oxygen Analyser Series PMA®

Portable version PMA 10
for mobile oxygen measurement

PMA 10

Application

Due to the extremely fast response time

of the M&C magneto-dynamic measuring

cell with no stagnant volume as well as the

negligible cross sensitivity from other sample

gas components, the portable M&C oxygen

analyser PMA 10 has a wide variety of applica-

tions. The analyser is a suitable and reliable

instrument for monitoring oxygen concentra-

tions in various gas analytical control applica-

tions including flue gas-, inert gas-, fermen-

tation processes-, food packing machines-,

ambient air- and laboratory process control

measurements.

Description

The portable M&C oxygen analyser PMA 10 is

an non-thermostated instrument which has

been designed for continuous and discontin-

uous measurement of oxygen concentrations

in dry and particle-free sample gases.
The PMA 10 is a reliable and easy-to-operate

instrument and immediately operable. The

light-weight instrument is built into a port-

able housing. The four measuring ranges

are displayed on the analogue meter with

30%/100% scale and the 100% O

2

range

also on the digital meter. One output signal

0-1V is available as standard. Sample gas

connections as well as a connector for the

output signal are located on the front panel

of the analyser. The connectors for incoming

power supply and optional O

2

alarm contact

are located at the rear panel. The sample gas

enters the analyser via an internal protec-

tive fine-filter. The required flow rate can be

adjusted at the flowmeter with needle valve,

mounted on the front panel upstream the

M&C measuring cell.
Options: mA-output signal, O

2

alarm,

rechargeable battery, internal pump and TÜV

certificate.

Measuring principle of M&C ox-

ygen analyser

The PMA 10 utilises the paramagnetic

principle of operation to measure oxygen

concentrations. The analyser measures the

paramagnetic susceptibility of the oxygen

in the sample gas by means of the M&C

magneto-dynamic measuring cell. The physi-

cal property which distinguishes oxygen from

other gases is its paramagnetism. It is signifi-

cantly higher comparing to other common

gases. This operation principle is one of the

most accurate and reliable procedures to

determine the oxygen concentration in a

gas mixture from 0 to 100 Vol.%. The robust

M&C cross-flow cell has no stagnant volume.

Advantages are the fast response time, the

flow rate up to 60 l/hr, the small volume of 2

ml, the extremely low drift, the absolute linear-

ity and the negligible cross sensitivity against

other sample gas components. With a proper

sample conditioning and pressure, the M&C

cell will never need replacing. The dumbbell

with a small mirror at its centre is mounted in

a strong inhomogenous magnetic field. The

paramagnetic oxygen strengthens the forces

on the diamagnetic dumbbell and causes a

shifting which is detected by a system con-

sisting of light beam, mirror and a photo cell.

A compensation current is induced via the

feedback coil on the dumbbell and leads to

a reset of the dumbbell into its zero-position.

The required current is linearly proportional to

the oxygen concentration.

Technical specifications and illustrations are without

obligation, subject to modifications. 05.97/08.09