Technical data dimensions – M&C TechGroup MP47-Z-BPR150 Data sheet User Manual
Page 2
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
10.4
Technical specifications and illustrations are without
obligation, subject to modifications. 01.08
Pump head with integrated bypass admission pressure regulator
Diaphragm pump
MP47-Z-BPR150
-Viton, 230V
MP47-Z-BPR150
-Viton, 115V
MP47-Z-BPR150
-Kalrez, 230V
MP47-Z-BPR150
-Kalrez, 115V
Part No.
02 P 1160
02 P 1160a
02 P 1165
02 P 1165a
Voltage
230V 50Hz
115V 60Hz
230V 50Hz
115V 60Hz
Material of sample
contacting parts
PTFE, PVDF, Viton
PTFE, PVDF, Kalrez
Protection class
IP 20 - DIN 40050
Pump capacity max.
6,6 l/min* in the range of adjustment (outlet pressure)
Operating pressure max.*
1,1 bar abs.
Operating pressure min.**
0,46 bar abs. for 250Nl/h / 0,62 bar abs. for 400Nl/h (outlet pressure 1,20 bar abs.)
Outlet pressure
1,15 – 1,3 bar abs. / at works 1,20 bar abs.
Gas temperature
-30 to +80°C
Ambient temperature
-10 to +40°C
Storage temperature
-15 to +60°C
Power consumption
110W
Current consumption
0,75A
1,7A
0,75A
1,7A
Gas connections
G1/4“ i DIN ISO 228/1
Electrical standard
EN 61010 part 1
Weight
4,6 Kg
*
Maximum operating pressure for a constant outlet pressure and as a result a constant flow rate of the measuring gas.
** The minimum operating pressure that still guarantees a constant pump capacity is changing according to the necessary gas volume flow adjusted behind the pump and to the outlet
pressure adjusted on the pressure regulator. The lower the necessary gas volume flow and the outlet pressure, the lower is he minimal possible operating pressure that still guarantees
a constant outlet pressure and therefore a constant sample gas volume flow.
Technical Data
Dimensions
Bypass admission
pressure regulator
Sample gas
IN OUT
66
18
80
110
160
G1/4"
4xM6 10 tief (deep)
34
47
247