Embracing Challenge

ƒ

ƒ

With integrated bypass admission pres-
sure regulator

ƒ

ƒ

Corrosion resistant

ƒ

ƒ

100% lubricant free

ƒ

ƒ

Gas tight

ƒ

ƒ

Maintenance free

ƒ

ƒ

Long service life

Special Features

Analytical Diaphragm Pump
Series MP®

Version MP47-Z-BPR150
with integrated bypass admission pressure regulator

MP47-Z-BPR150

Application

The diaphragm pump MP47-Z-BPR150 is suit-

able for the 100 % oil-free delivery of corrosive

gases with changing pressure conditions. The

capacity and construction of the pump is

especially designed for the problems in the

analyze technique. The pump is gas-tight and

operates without any maintenance.

Description

All sample contacting parts of the diaphragm

pump MP47-Z-BPR150 are made of PTFE,

PVDF and Viton or Kalrez. The lifted gas

remains analytically pure due to the fact that

the pump works absolutely without any lubri-

cant. A special diaphragm and valve system

guaranties a long life time and no mainte-

nance is necessary. The pump is available for

230V or 115V.
One problem in the analyze technique, when

lifting gases via filters, is that the filters are

clogged after a longer period of time. This

causes an increasing differential pressure

above the filter and subsequently, a higher

suction pressure for the pump is necessary.

This, however, results in a decrease of the

adjusted pump capacity.
In order to avoid this negative influence, the

M&C diaphragm pump MP47-Z-BPR150 is

equipped with a fully adequate bypass admis-

sion pressure regulator inside the pump head.

Independent from the changing entry pres-

sure, the bypass admission pressure regula-

tor keeps the pressure at the pump outlet

constant, so that the adjusted pump capacity

resulting thereof remains constant as well.
For example, the entry pressure may vary

from 0,46 bar abs. to 1,1 bar abs. for a pump

capacity of 250Nl/h.
The outlet pressure is adjusted at works to

200mbar overpressure (adjustable up to

300mbar).

Technical specifications and illustrations are without