beautypg.com

M&C TechGroup BA-C Data sheet User Manual

Air clean up supply series ba, Version ba-c, Special features

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

12.2

Embracing Challenge

ƒ

ƒ

High degree of purity:
C

n

H

m

< 0,1ppm,

H

2

O dew point < -100 °F

CO

2

< 2ppm

ƒ

ƒ

Long service life

ƒ

ƒ

Minimum operating costs

ƒ

ƒ

Status self-monitoring

ƒ

ƒ

Compact, easy-to-use and easy-to-
service 19 inch plug-in or wall mounting
housing

ƒ

ƒ

Safe handling and easy servicing

Special Features

Air clean up supply series BA

Version BA-C

BA-C

Application

Dilution extractive technology requires a con-

stant low dew point clean air supply that

removes the concentrations well below the

lower detection limits (LDL) of the analyzers

used. This translates to low ppb concentra-

tions in some cases.
With ever decreasing concentrations of mea-

sured pollutants, an increased purity specifi-

cation is required to maintain system repea-

tability since simply changing dilution ratios

can cause serious complications
The air clean up supply BA-C has been deve-

loped in order to achieve proper dilution air

for gas sample dilution probes (see data sheet

2-1.1.7), independent of gas cylinders.
The BA-C unit can also be used as a zero gas

generator when calibrating I.R.-analysers for

immission monitoring and for hydrocarbon

free burner air when measuring hydrocarbons

with a FID, FTIR and gas Chromatography also

need an ultra-pure reference as well as carrier

air supplies.

Description

The M&C-air clean up supply version BA-C has

been designed as compact, easy-to-use and

easy-to-service 19 inch and wall mounting

module.
With version BA-C compressed air is taken via

the rear-mounted air filter and is directed to

particle and condensate separation by means

of a high-performance filter (0.01 micron of

filter fineness) with an integrated, floating

condensate drainer.
There upon the compressed air is cleaned as

a result of low power catalytic oxidation by

means of a platinum/palladium catalyst of

hydrocarbon traces
(C

n

H

m

-> CO

2

+ 2H

2

O).

Drying the condensate-free compressed air

to <-100 °F dew point and removing NO

x

and

CO

2

takes place in two mol sieve columns.

Technical specifications and illustrations are without

obligation, subject to modifications. 09.96/01.07