Air clean up supply series ba, Version ba-c, Special features
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
12.2
Embracing Challenge
High degree of purity:
C
n
H
m
< 0,1ppm,
H
2
O dew point < -100 °F
CO
2
< 2ppm
Long service life
Minimum operating costs
Status self-monitoring
Compact, easy-to-use and easy-to-
service 19 inch plug-in or wall mounting
housing
Safe handling and easy servicing
Special Features
Air clean up supply series BA
Version BA-C
BA-C
Application
Dilution extractive technology requires a con-
stant low dew point clean air supply that
removes the concentrations well below the
lower detection limits (LDL) of the analyzers
used. This translates to low ppb concentra-
tions in some cases.
With ever decreasing concentrations of mea-
sured pollutants, an increased purity specifi-
cation is required to maintain system repea-
tability since simply changing dilution ratios
can cause serious complications
The air clean up supply BA-C has been deve-
loped in order to achieve proper dilution air
for gas sample dilution probes (see data sheet
2-1.1.7), independent of gas cylinders.
The BA-C unit can also be used as a zero gas
generator when calibrating I.R.-analysers for
immission monitoring and for hydrocarbon
free burner air when measuring hydrocarbons
with a FID, FTIR and gas Chromatography also
need an ultra-pure reference as well as carrier
air supplies.
Description
The M&C-air clean up supply version BA-C has
been designed as compact, easy-to-use and
easy-to-service 19 inch and wall mounting
module.
With version BA-C compressed air is taken via
the rear-mounted air filter and is directed to
particle and condensate separation by means
of a high-performance filter (0.01 micron of
filter fineness) with an integrated, floating
condensate drainer.
There upon the compressed air is cleaned as
a result of low power catalytic oxidation by
means of a platinum/palladium catalyst of
hydrocarbon traces
(C
n
H
m
-> CO
2
+ 2H
2
O).
Drying the condensate-free compressed air
to <-100 °F dew point and removing NO
x
and
CO
2
takes place in two mol sieve columns.
Technical specifications and illustrations are without