Technicaldata – M&C TechGroup PMA 30A_D Data sheet User Manual
Page 3
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
13.4
Technical specifications and illustrations are without
obligation, subject to modifications. 05.97/11.12
TechnicalData
WARNING!IMPORTANT!
An external fine filter must always be used at the gas inlet of the analyser. Depending
on the composition of the sample gas, it may be necessary to use a sample conditio-
ning system. Without precautions, the analyser is only suitable for measuring of non-
hazardous gases or gasmixtures in non-hazardous areas.
VersionPMA30thermostatedoxygenanalyserincompact19-inchcassette
Part No.
PMA 30/D: 03A2005 = 230V 50Hz, 03A2005a = 115V 60Hz
PMA 30A/D: 03A2001 = 230V 50Hz, 03A2005a = 115V 60Hz
Measuring ranges
selectable for 0-1, 0-3, 0-10, 0-30 und 0-100 Vol.% O
2
, linear
Expanded measuring ranges***
option: zero suppression for spreading of measuring range, adjusted by the standard
ranges, for example 20-21% (1%), 18-21% (3%), 90-100% (10%), max.97-100%, incl. process pressure compensation
type SDPD** Part No.: 03A9535
External measuring range indication
potential free contacts for each selected range, contact rating max. 48V DC, 0,5A
Remote range control
Remote range control for all 5 ranges by means of external potential free contacts
Indication
PMA 30A/D: analogue meter with a scale of 0-30 and 0-100% for each selected range,
digital meter, 4
1
/
2
digit 9 mm high LCD-indicator for 0-100% O
2
reading, selectivity 0,1vol.% O
2
PMA 30/D: digital meter 4
1
/
2
digit 18 mm high LCD for 0-100% O
2
reading, selectivity 0,1 vol.% O
2
Output signals
0-10V
DC
non-isolated, load > 100 KΩ, for the range of 100 vol.%, and
0-20mA* or 4-20mA* for the chosen range, non-isolated, max. load 300Ω / 900Ω
Response time for 90% FSD
< 3 seconds at 60 Nl/hr air
Accuracy after calibration
deviation:
analogue signal output = ±1% of span at range 3-100% / digital indicator = ±0,1 vol.% O
2
= ±2% of span at range 1%
Reproducibility
deviation: analogue = < 1% of span / digital = ± 0,1 vol.% O
2
Influence of ambient temperature
no influence up to 45 °C
Influence of barometric pressure
the oxygen reading varies in direct proportion to changes of the barometric pressure
Influence of sample gas flow
variation in gas flow between 0-60 Nl/hr air will cause a difference of < 0,1 vol.% O
2
Sample gas inlet pressure
0,01 up to 1 bar g, (PMA 30 required admission pressure for competent flow rate, no pump inside)
Sample gas outlet pressure
outlet of analyser must discharge freely into atmosphere;
option: process pressure compensation PD** in a range of 0,4 to 1,6 bar absolut Part No.: 03A9530
Flow rate of sample gas
max. 60 Nl/hr air, adjustable with needle valve on the flowmeter 7-70 Nl/hr
Temperature of sample gas
-10 °C up to +40 °C dry gas
O
2
-transducer temperature
fixed at +50 °C
Ambient temperature
-10 °C up to +45°C
Storage temperature
-20 °C up to +60 °C, relative humidity 0-90% RH
Power supply
internal power unit for 230V
AC
standard or 115V
AC
available (a)* +/-15%, 40-60Hz, 35,5VA
Electrical connections
mains supply: 3-pole chassis plug with 2 m of cable; signals: 9 and 25-pole plugs
Materials in contact with sample gas
Platinum, Glass, PTFE, PVDF, Stainless Steel 316Ti, FPM, Epoxy resin
Sample gas connections
1/8" NPT i, optional hose connector DN 4/6 available
Flow alarm
thermo-conductive flow sensor downstream mounted after measuring cell
Status alarm
for low sample flow, transducer temperature < 40 °C, light beam defect, power failure:
LED indication and potential free change over contact, contact rating max. 48V
DC
, 0,5A (2x)
Protection / electrical standard
IP 20 EN 60529 / EN 61010
Housing / front colour
3U/28HP - 19-inch cassette European standard / silver
Dimension / weight
height 3U, width 28HP, length 230 mm + approx. 60 mm installation space / approx. 2,5 kg
Options
03A9150
Wall mountage housing out of aluminium type 30-19 with 19“-insertion system (3U/42TE), incl. clamp box, clamp
boards and filter, with window. protection class: IP65.
*
Please specify with order.
**
Without SIL-certification
*** Standard with analogue and digital meter.