beautypg.com

M&C TechGroup FP-200P190-D Data sheet User Manual

Page 2

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

7.12

Technical specifications and illustrations are without

obligation, subject to modifications. 09.96/11.06

Flow rate in dependence of inlet pressure

Dimensions

Filter FP-70P200-D

Filtration in a setting area

Dimensions in mm

13

G3/8"

G1/4"

84

54

58

42

sample in

bypass

sample

ø 5

190

290

60

out

connection GL25

13

connection GL25

sample

out

bypass

5

42

58

G3/8"

G1/4"

60

290

190

sample in

Filter FP-70-P190-D / FP200P190-D

Filtration with self-cleaning effect

through bypass system

Technical Data

l/h

r

0 bar

0,2 bar

0,5 bar

1,0 bar

1,5 bar

2,0 bar

0

200

400

600

800

1000

1200

1400

1600

Bypass

Sample

Liquid bypass filter series FP®

FP-70P200-D

FP-70P190-D

FP-200P190-D

Part No.

02F1400

02F1440

02F1450

Filtration in a settling area

yes

no

Filtration with self-cleaning-effect

no

yes

Filter element fineness

70 µm

200 µm

Bypass-connections

G3/8" i DIN ISO 228/1 and GL25-12 GL25-12 and G3/8" i DIN ISO 228/1

Sample connection

G1/4" i DIN ISO 228/1

Max. pressure at 20 °C

max. 2 bar abs. or max. 5 bar abs. with GL-connecting adapter

Bypass flow max.

1500 l/hr

Sample max.

180 l/hr

Sample temperature max.

80 °C

Ambient temperature max.

80 °C

Filter surface

180 cm

2

Volume

100 ml

Weight

approx. 650 g

Method of mounting

wall mounting

Material of sample contacting parts

PVDF, PP, glass, FPM