M&C TechGroup PSS-10_1 Data sheet User Manual
Portable gas conditioning unit series pss, Special features, Application
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
14.6
Embracing Challenge
Low maintenance and self-monitoring
Outlet dew point adjustable from +2 °C
to +15 °C
Dew point stability ± 0,1 °C
Operational in 20 minutes
Compact construction
Light weight
Optimum reliability
Jet Stream heat exchangers in 3 stand-
ard materials
Universal equipment possible
Special Features
Portable Gas Conditioning Unit
series PSS®
Version PSS-10/1 for 480 Nl/hr gas flow rate
PSS-10/1
Application
The portable gas conditioning unit PSS-10/1
is designed to carry out precise gas analyses
at different locations. The entire gas condi-
tioning unit is housed in a robust aluminium-
frame case to enable you to carry out your gas
analysis rapidly, safely and with little mainte-
nance work. The PSS-10/1 gas conditioning
unit is suitable for variable, discontinuous use
as for continuous operation. The components
built into the PSS-10/1 gas conditioning unit
are envisaged for “standard use“. Obviously
our extensive Product Range provides addi-
tional components.
Description
The M&C PSS-10/1 gas conditioning unit is
fitted with an two stage ECP20-2 gas cooler
which cools the sample gas to a constant +
5°C irrespective of the ambient temperature.
As soon as the operating temperature of
<+8°C is reached after start up, the gas pump
N9KPE is switched on automatically by the gas
cooler status contact. The SR 25.1 peristaltic
pumps ensures a constant condensate drain,
making long-term measurement possi-
ble without problems. The corresponding
particle filtration is carried out by the FP-2T
micron filter. This all makes the portable gas
conditioning unit a complete gas condition-
ing system for most gas-analysis devices.
Technical specifications and illustrations are without
obligation, subject to modifications. 10.96/12.11