Application

The M&C liquid-stop LB-1.. is suitable for

protecting analysers against inrush of liquids

out of the gas conditioning unit which is

mounted before the analyser. This avoids

major damages of the analyser.
The most useful position of the liquid stop

LB-1.. is behind the gas conditioning unit,

directly before the flowmeter of the analy-

ser.

Liquid stop LB-PV and LB-SS

for protecting gas analysers and gas-phase

chromatographs against liquids

Special Features
Sure protection against liquid inrush
Reliable removal of liquid coming

through
Also suitable for high pressures
Available in stainless steel and PVDF
Easy change of the hydrophobic

protective membrane
With wall mounting holder

Description

The hydrophobic protective membrane of

the liquid-stop LB-1.. is placed between the

two parts of the housing which are screwed

together. It is lined with a porous glass filter

frit in order to secure stable proportions.
The pore sizes of the protective membrane

are designed in such a way that gas mole-

cules and steam can pass through but liquid

molecules are retained.
The gas inlet and outlet are positioned in

an horizontal line inside the housing. The

liquid outlet shows downwards when being

mounted.
Due to the horizontal flow direction of the

gases and because possible liquids are drai-

ning off on the protective membrane due to

its gravity, the leaking through of liquids to

the analyser is avoided.
Eventually occuring liquid can be let off via

a peristaltic pump SR25.1, separator and

automatic liquid drain ADS-SS or a conden-

sate vessel TG1 (dip tank). The LB-1 has got a

respective connection possibility.
Changing the membrane is easy to handle.

The optimum positioning of the sealing

O-rings guarantees always a sure sealing of

both housing parts.
The filter inlet and outlet can be turned by

180° on the wall mounting holder so that a

flexible adaption to local circumstances is

possible during mountage.

LB-PV, LB-SS