8

Gas sampling and gas conditioning technology

2-1.1.6-ME

7

TECHNICAL DATA

Mains supply

230V 50/60Hz 800W*, optional 115V 60Hz 800W

Electrical connection

terminals; max 4mm², 2x PG13,5 cable gland

Protective type

IP54 (EN 60529)*

Ambient temperature

thermostat:
-20°C to 60°C
PT100, thermocouple:
-20°C to 80°C

Operating temperature

0-180°C*

Ready for operation

after 2 hours

Temperature status alarm point

30°C to T

Set

*

Alarm contact rating

250V 3A AC, 0,25A DC

Connection sample outlet

1/4"-NPT i., for max. 10mm tube connectors

Connection backflush

tube fitting 8mm ø

Connection calibration gas

tube 6mm ø

Filter chamber volume

120 ml

Sample pressure

0,4 to 6 bar abs.

Material of sample conducting parts

SS316Ti, Viton*
alternative = SS316Ti, PTFE

* = Standard

Differential pressure and T

90

-time at different flow rates:

P and T

90

at flow of

100

200

500

1000

1500

Nl/hr

P with new filter element

S-2K 150 or F-0,1GF150:

0,007

0,011

0,020

0,035

0,04

bar

T

90

-time without sample tube/pre filter:

6,0

3,5

1,0

<0,5

<0,5

sec.



8

APPLICATIONS

The SP2200... gas sample probes are used for continuous extraction of gases from dust loaded,
high-temperature or humid processes.
A host of different variations in sampling technology as well as some built-in accessories offer a wide
range of possible applications for nearly all problems encountered in taking gas samples from proc-
esses with high dust content where an effective backflush is necessary. Sampling from processes up
to a maximum of 1900°C are possible with this probe.