Receipt of goods, Preparation and installation, 12 receipt of goods – M&C TechGroup PSP4000-H_C_T Operator's manual User Manual
Page 11: 13 preparation and installation
11
Gas sampling and gas conditioning technology
2-2.1-ME
The heated sample line is rapidly and safely mechanically coupled with a tightening bracket including
integrated quick-coupling.
To ensure simple filter changes, the complete filter unit including the connected sample line can be
drawn out of the heat-insulated protective casing.
In the PSP400H/C special version, calibration-gas into the probe is possible without disassembling
the probe via a special threaded connecting joint.
The special version PSP4000H/C/T allows the process temperature to be measured by way of a
thermo-element which is introduced into this threaded connecting joint. Its length can be adjusted up
to the end of the sampling tube. The standard length of this thermocouple sensor is 600 mm.
Optionally available is the PSP transport case for transporting the PSP4000-H... gas-sample probe
including a maximum 5 m heated sample line PSP 4M4/6 resp. max. 8m for type PSP-W4M4/6.
12
RECEIPT OF GOODS
The gas sample probe PSP4000-H... is normally delivered in two packaging units:
1. Basic gas-sample probe with optional mounting accessories.
2. Optional sample tubes.
The gas sample probe should be removed carefully from the packaging and checked immediately for
completeness against the delivery note.
Check the goods for any damage incurred during transport and if necessary inform your transport
insurer of any damage.
13
PREPARATION AND INSTALLATION
Select the optimal sampling point in accordance with the generally applicable guidelines or consult the
competent persons.
The sampling point should be easily accessible.
Locate the probe connections in such a way that the connections' temperature is always above the
acid dew point in order to avoid corrosion and blockage problems.
If the ambient temperature in the area of the connections is >80°C as a result of radiated heat, then a
radiated-heat deflector must be mounted to protect the probe.
The probe can also be mounted with an optional mounting flange or R2" threaded mounting fitting.
Before mounting, the probe must be adjusted to the existing operating conditions.