Using the efi densitometer ed-100, Setting up the ed-100 – Toshiba GA-1121 User Manual
Page 50
2
2-14
Color Calibration
14.
through
until all the strips have been measured.
N
OTE
:
You must take the measurements in the order outlined in the Status field.
15.
When all the patches have been read successfully, click Accept.
16.
Click Apply in the Calibrator window to implement the new calibration measurement.
17.
Click OK in the Information dialog box.
This completes the GA-1121 calibration process.
Using the EFI Densitometer ED-100
The EFI Densitometer ED-100 is a hand-held instrument that measures the density of
printed ink and toner. Like the DTP32, the ED-100 is a reflection
densitometer
that
you can use to calibrate the GA-1121.
Setting up the ED-100
To use the ED-100 for calibration purposes, you must first set it up for use with your
computer. For complete instructions on setting up the ED-100, see the documentation
that is included with the instrument.
The following illustration shows the components of the ED-100.
Indicator light
Sample aperture
(underneath)
Measure button