Calibrating the dtp32 – Toshiba GA-1121 User Manual
Page 108
A
A-10
Color Measurement Instruments
4.
Plug the small connector on the adapter cable into the side of the X-Rite DTP32 and
plug the AC adapter into a wall outlet.
5.
Turn on the computer.
6.
Calibrate the DTP32 (see the following section).
7.
Calibrate the GA-1121 using ColorWise Pro Tools and the DTP32 (see
).
Calibrating the DTP32
You will need the black-and-white X-Rite Auto-Cal Strip included with the
densitometer. Calibrating the densitometer does not require ColorWise Pro Tools.
T
O
CALIBRATE
THE
X-R
ITE
DTP32
1.
Connect the densitometer to the computer and supply power (see
2.
From the Main Menu on the X-Rite DTP32 display, press the p1 key once to reach p2.
3.
Press the cal key.
Calibrating motor speed
is displayed, followed by the words
INSERT CAL STRIP
.
4.
Insert the X-Rite Auto-Cal Strip into the 35mm slot on the front of the X-Rite DTP32—
arrow end first—until it stops or the densitometer starts pulling the strip.
Reading
appears momentarily, followed by the density values and
CALIBRATION OK
.
The densitometer automatically returns to the
MAIN MENU
.
If
UNRECOGNIZABLE STRIP
appears, repeat the process or try cleaning the strip (see the
X-Rite DTP32 Operating Manual).
AC adapter
Small connector