Calibrating the dtp32 – Toshiba GA-1121 User Manual

Page 108

background image

A

A-10

Color Measurement Instruments

4.

Plug the small connector on the adapter cable into the side of the X-Rite DTP32 and
plug the AC adapter into a wall outlet.

5.

Turn on the computer.

6.

Calibrate the DTP32 (see the following section).

7.

Calibrate the GA-1121 using ColorWise Pro Tools and the DTP32 (see

page 2-22

).

Calibrating the DTP32

You will need the black-and-white X-Rite Auto-Cal Strip included with the
densitometer. Calibrating the densitometer does not require ColorWise Pro Tools.

T

O

CALIBRATE

THE

X-R

ITE

DTP32

1.

Connect the densitometer to the computer and supply power (see

page A-8

).

2.

From the Main Menu on the X-Rite DTP32 display, press the p1 key once to reach p2.

3.

Press the cal key.

Calibrating motor speed

is displayed, followed by the words

INSERT CAL STRIP

.

4.

Insert the X-Rite Auto-Cal Strip into the 35mm slot on the front of the X-Rite DTP32—
arrow end first—until it stops or the densitometer starts pulling the strip.

Reading

appears momentarily, followed by the density values and

CALIBRATION OK

.

The densitometer automatically returns to the

MAIN MENU

.

If

UNRECOGNIZABLE STRIP

appears, repeat the process or try cleaning the strip (see the

X-Rite DTP32 Operating Manual).