Calibrating the ga-1121 with the es-1000 – Toshiba GA-1121 User Manual
Page 47
2
2-11
Using the EFI Spectrometer ES-1000
The following illustration shows the different parts of the ES-1000.
Calibrating the GA-1121 with the ES-1000
You can use ColorWise Pro Tools in conjunction with the ES-1000 to calibrate
the GA-1121.
T
O
CALIBRATE
THE
GA-1121
USING
THE
ES-1000
1.
Start Calibrator.
For instructions on starting Calibrator, see
2.
Choose EFI Spectrometer ES-1000 as the measurement method.
3.
In the Check Print Settings pane, choose the desired calibration set.
Choose the appropriate calibration set for the mode and type of media which are
needed for calibration.
N
OTE
:
For this calibration to take effect, the calibration set must be associated with one
or more output profiles. There is no need to make any new associations for the default
calibration set, since it is already associated with the default output profile.
4.
In the Generate Measurement Page pane, click Print.
The Print Options dialog box appears.
Sample aperture
Measure button
White tile reference
Calibration cradle