Appendix a specifications, Analog input, Input characteristics – National Instruments Low-Cost Multifunction I/O Board for ISA Lab-PC+ User Manual
Page 82: Transfer characteristics
© National Instruments Corporation
A-1
Lab-PC+ User Manual
Appendix A
Specifications
This appendix lists the specifications of the Lab-PC+. These specifications are typical at 25
°
C
unless otherwise stated. The operating temperature range is 0
°
to 70
°
C.
Analog Input
Input Characteristics
Number of channels ......................................
8 single-ended, 4 differential
Type of ADC.................................................
Successive
approximation
Resolution .....................................................
12 bits, 1 in 4,096
Maximum sampling rate ...............................
83
ksamples/s
Input signal ranges ........................................
Board Gain
(Software
Board Range
(Jumper Selectable)
Selectable)
bipolar
unipolar
1
±
5 V
0 to 10 V
2
±
2.5 V
0 to 5 V
5
±
1 V
0 to 2 V
10
±
0.5 V
0 to 1 V
20
±
0.25 V
0 to 0.5 V
50
±
0.1 V
0 to 0.2 V
100
±
0.05 V
0 to 0.1 V
Input coupling ...............................................
DC
Overvoltage protection..................................
±
45 V powered on,
±
45 V powered off
Inputs protected .......................................
ACH<0..7>
FIFO buffer size ............................................
512
samples
Data transfers ................................................
DMA, interrupts, programmed I/O
DMA modes ..................................................
Single
transfer
Transfer Characteristics
Relative accuracy ..........................................
±
1.0 LSB typ,
±
1.5 LSB max
DNL ..............................................................
±
0.5 LSB typ,
±
1 LSB max
No missing codes ..........................................
12 bits, guaranteed
Offset error
Pregain error after calibration .................
Adjustable to 0 V
Postgain error after calibration................
Adjustable to 0 V
Gain error (relative to calibration reference)
After calibration ......................................
Adjustable to 0%
Before calibration....................................
±
0.76% of reading (7,600 ppm) max
Gain
≠
1 with gain error
adjusted to 0 at gain = 1 ........................
±
0.5% of reading (500 ppm) max