beautypg.com

Appendix a specifications, Analog input, Input characteristics – National Instruments Low-Cost Multifunction I/O Board for ISA Lab-PC+ User Manual

Page 82: Transfer characteristics

background image

© National Instruments Corporation

A-1

Lab-PC+ User Manual

Appendix A
Specifications

This appendix lists the specifications of the Lab-PC+. These specifications are typical at 25

°

C

unless otherwise stated. The operating temperature range is 0

°

to 70

°

C.

Analog Input

Input Characteristics
Number of channels ......................................

8 single-ended, 4 differential

Type of ADC.................................................

Successive

approximation

Resolution .....................................................

12 bits, 1 in 4,096

Maximum sampling rate ...............................

83

ksamples/s

Input signal ranges ........................................

Board Gain

(Software

Board Range

(Jumper Selectable)

Selectable)

bipolar

unipolar

1

±

5 V

0 to 10 V

2

±

2.5 V

0 to 5 V

5

±

1 V

0 to 2 V

10

±

0.5 V

0 to 1 V

20

±

0.25 V

0 to 0.5 V

50

±

0.1 V

0 to 0.2 V

100

±

0.05 V

0 to 0.1 V

Input coupling ...............................................

DC

Overvoltage protection..................................

±

45 V powered on,

±

45 V powered off

Inputs protected .......................................

ACH<0..7>

FIFO buffer size ............................................

512

samples

Data transfers ................................................

DMA, interrupts, programmed I/O

DMA modes ..................................................

Single

transfer

Transfer Characteristics
Relative accuracy ..........................................

±

1.0 LSB typ,

±

1.5 LSB max

DNL ..............................................................

±

0.5 LSB typ,

±

1 LSB max

No missing codes ..........................................

12 bits, guaranteed

Offset error

Pregain error after calibration .................

Adjustable to 0 V

Postgain error after calibration................

Adjustable to 0 V

Gain error (relative to calibration reference)

After calibration ......................................

Adjustable to 0%

Before calibration....................................

±

0.76% of reading (7,600 ppm) max

Gain

1 with gain error

adjusted to 0 at gain = 1 ........................

±

0.5% of reading (500 ppm) max