beautypg.com

National Instruments Low-Cost Multifunction I/O Board for ISA Lab-PC+ User Manual

Page 80

background image

Chapter 5

Calibration

© National Instruments Corporation

5-7

Lab-PC+ User Manual

1. Adjust the Analog Output Offset

Adjust the analog output offset by measuring the output voltage generated with the DAC set
at negative full-scale (0). This output voltage should be V

-fs

±

0.5 LSB. For bipolar output,

V

-fs

= -5 V, and 0.5 LSB = 1.22 mV.

For analog output Channel 0:

a. Connect the voltmeter between DAC0 OUT (pin 10 on the I/O connector) and AGND

(pin 11).

b. Set the analog output channel to -5 V by writing -2,048 to the DAC.

c. Adjust trimpot R2 until the output voltage read is -5 V.

For analog output Channel 1:

a. Connect the voltmeter between DAC1 OUT (pin 12 on the I/O connector) and AGND

(pin 11).

b. Set the analog output channel to -5 V by writing -2,048 to the DAC.

c. Adjust trimpot R4 until the output voltage read is -5 V.

2. Adjust the Analog Output Gain

Adjust the analog output gain by measuring the output voltage generated with the DAC set at
positive full-scale (4,095). This output voltage should be V

+fs

±

0.5 LSB. For bipolar output,

V

+fs

= +4.99756 V, and 0.5 LSB = 1.22 mV.

For analog output Channel 0:

a. Connect the voltmeter between DAC0 OUT (pin 10 on the I/O connector) and AGND

(pin 11).

b. Set the analog output channel to +4.99756 V by writing 2,047 to the DAC.

c. Adjust trimpot R1 until the output voltage read is +4.99756 V.

For analog output Channel 1:

a. Connect the voltmeter between DAC1 OUT (pin 12 on the I/O connector) and AGND

(pin 11).

b. Set the analog output channel to +4.99756 V by writing 2,047 to the DAC.