Page 98: 5 route:scan? 92, 6 route:scan 92, 7 route:scan:mode? [min|max|def] 92, Route:close

8.4.3.2
ROUTe:CLOSe
Enables the specified channel.
ROUT:CLOS 3
The
does not apply to memory channels.
8.4.3.3
ROUTe:OPEN?
Returns the state of the specified channel.
Example: ROUT:OPEN? 1
Response: 0
The
enabled this query returns 0. If the specified channel is disabled, this query re-
turns 1.
8.4.3.4
ROUTe:OPEN
Disables the specified channel.
Example: ROUT:OPEN 4
The
does not apply to memory channels.
8.4.3.5
ROUTe:SCAN?
Returns a comma delimited list of the enabled channels.
ROUT:SCAN?
(@2,3,4)
The list of channels is enclosed in parenthesis and preceded with the @
symbol.
8.4.3.6
ROUTe:SCAN
Sets the channels for scanning.
ROUT:SCAN 2,3
The
enable. Channels that are omitted from this list are disabled. If the
parameter is omitted, all channels are disabled. The *RST command sets chan-
nel 1 ON, and all other channels OFF.
8.4.3.7
ROUTe:SCAN:MODE? [MIN|MAX|DEF]
Returns the current scan mode setting.
1529 Chub-E4 Thermometer Readout