7 system:time? 116, 8 system:time ,, 116, 11 password commands 116 – Fluke THE HART 1529 User Manual
Page 122: 1 system:password:cdisable 116, 2 system:password:cenable 116, 3 system:password:cenable:state? 116, System:time, System:time ,,

8.4.10.7
SYSTem:TIME?
Returns the current system time.
SYST:TIME?
11,43,23
The response is returned in the format,
Note: The system time is always returned using 24-hour format regardless of
the current time format setting.
8.4.10.8
SYSTem:TIME
Sets the current system time to the specified hour, minute, and second.
SYST:TIME 11,43,23
The
eter is a one or two-digit number (0 to 59). The
two-digit number (0 to 59).
Note: The system time is always set using 24-hour format regardless of the cur-
rent time setting.
8.4.11
Password Commands
The password commands are used for enabling and disabling the password-pro-
tected commands, for querying and setting the probe password-protected state,
and for setting a new password.
8.4.11.1
SYSTem:PASSword:CDISable
Disables password protected commands.
SYST:PASS:CDIS
Sending this command prevents the user from accessing the password-pro-
tected settings.
8.4.11.2
SYSTem:PASSword:CENable
Enables password protected commands.
SYST:PASS:CEN 1234
The
sent to allow the user to access password-protected settings. After sending this
command, use the SYST:PASS:CEN:STAT? command to verify success.
8.4.11.3
SYSTem:PASSword:CENable:STATe?
Returns the current password protection state.
1529 Chub-E4 Thermometer Readout
User’s Guide