System:communicate:serial:feed? [min|max|def, System:communicate:serial:feed |min|max|def, System:communicate:serial:linefeed? [min|max|def – Fluke THE HART 1529 User Manual
Page 118

1
A value of 1 is returned if serial port echo (duplex) is ON. A value of 0 is re-
turned if serial port echo (duplex) is OFF. Specifying the MIN parameter re-
turns a value of 0. Specifying the MAX parameter returns a value of 1.
Specifying the DEF parameter returns a value of 1.
8.4.9.4
SYSTem:COMMunicate:SERial:FDUPlex
Sets the serial port echo (duplex) setting.
SYST:COMM:SER:FDUP MAX
The
(0 or OFF). Specifying the MIN, MAX, or DEF parameter sets the serial port
echo (duplex) to the minimum, maximum, or default respectively.
8.4.9.5
SYSTem:COMMunicate:SERial:FEED? [MIN|MAX|DEF]
Returns the state of the serial port printing.
SYST:COMM:SER:FEED?
0
A value of 1 is returned if serial port printing is enabled. A value of 0 if re-
turned is serial port printing is disabled. Specifying the MIN parameter returns
a value of 0. Specifying the MAX parameter returns a value of 1. Specifying
the DEF parameter returns a value of 0.
8.4.9.6
SYSTem:COMMunicate:SERial:FEED
Sets the state of the serial port printing.
SYST:COMM:SER:FEED DEF
The
Specifying the MIN, MAX, or DEF parameter sets the serial port printing value
to the minimum, maximum, or default respectively. The *RST command sets
the serial port printing to OFF.
8.4.9.7
SYSTem:COMMunicate:SERial:LINefeed? [MIN|MAX|DEF]
Returns the state of the serial port linefeed setting.
SYST:COMM:SER:LIN?
1
A value of 1 is returned if linefeed is enabled. A value of 0 is returned if line-
feed is disabled. Specifying the MIN parameter returns a value of 0. Specifying
the MAX parameter returns a value of 1. Specifying the DEF parameter returns
a value of 1.
1529 Chub-E4 Thermometer Readout
User’s Guide
112