Calibrate:parameter:scale? [min|max|def – Fluke THE HART 1529 User Manual

Page 105

background image

8.4.5.7

CALibrate:PARameter:OFFSet |MIN|MAX|DEF

Sets the calibration offset parameter for the specified channel.

CAL1:PAR:OFFS2 5

The CALibrate suffix, , specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion offset suffix specifies the resistance range parameter (in ohms) for
PRTs (1) or Thermistors (2). For Thermocouple channels, the calibration offset
suffix is ignored and may be omitted. The paramter specifies the
value for the specified offset parameter. Specifying the MIN, MAX or DEF pa-
rameter sets the offset parameter to the minimum, maximum, or default value
respectively.

This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN com-
mand (see Section 9.4.11, Password Commands).

8.4.5.8

CALibrate:PARameter:SCALe? [MIN|MAX|DEF]

Returns the calibration scale parameter for the specified channel.

CAL1:PAR:SCAL2?

0

The CALibrate suffix, , specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion scale suffix specifies the resistance range parameter for PRTs (1) or
Thermistors (2). For Thermocouple channels, the calibration scale suffix is
ignored and may be omitted. Note: For PRT/Thermistor channels, the value re-
turned by this command is always in ohms.

For PRT/Thermistor channels, specifying the MIN parameter returns a value of
-9 when the scale suffix is 1 (for PRTs) and -9000 when the scale suffix
is 2 (for Thermistors). For Thermocouple channels, specifying the MIN
parameter returns a value of -9. For PRT/Thermistor channels, specifying the
MAX parameter returns a value of 9 when the scale suffix is 1 (for PRTs)
and 9000 when the scale suffix is 2 (for Thermistors). For Thermocouple
channels, specifying the MAX parameter returns a value of 9. Specifying the
DEF parameter returns a value of 0.

8.4.5.9

CALibrate:PARameter:SCALe |MIN|MAX|DEF

Sets the calibration scale parameter for the specified channel.

CAL1:PAR:SCAL2 2

The CALibrate suffix, , specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion scale suffix specifies the resistance range parameter (in ohms) for
PRTs (1) or Thermistors (2). For Thermocouple channels, the calibration scale
suffix is ignored and may be omitted. The parameter specifies the

99