Calibrate:parameter:scale? [min|max|def – Fluke THE HART 1529 User Manual
Page 105

8.4.5.7
CALibrate
Sets the calibration offset parameter for the specified channel.
CAL1:PAR:OFFS2 5
The CALibrate suffix,
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion offset suffix
PRTs (1) or Thermistors (2). For Thermocouple channels, the calibration offset
suffix
value for the specified offset parameter. Specifying the MIN, MAX or DEF pa-
rameter sets the offset parameter to the minimum, maximum, or default value
respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN
mand (see Section 9.4.11, Password Commands).
8.4.5.8
CALibrate
Returns the calibration scale parameter for the specified channel.
CAL1:PAR:SCAL2?
0
The CALibrate suffix,
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion scale suffix
Thermistors (2). For Thermocouple channels, the calibration scale suffix
ignored and may be omitted. Note: For PRT/Thermistor channels, the value re-
turned by this command is always in ohms.
For PRT/Thermistor channels, specifying the MIN parameter returns a value of
-9 when the scale suffix
parameter returns a value of -9. For PRT/Thermistor channels, specifying the
MAX parameter returns a value of 9 when the scale suffix
and 9000 when the scale suffix
channels, specifying the MAX parameter returns a value of 9. Specifying the
DEF parameter returns a value of 0.
8.4.5.9
CALibrate
Sets the calibration scale parameter for the specified channel.
CAL1:PAR:SCAL2 2
The CALibrate suffix,
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion scale suffix
PRTs (1) or Thermistors (2). For Thermocouple channels, the calibration scale
suffix
99
8 Digital Communications Interface
Commands