3 display:lamp? [min|max|def] 102, 4 display:lamp ||min|max|def 102, 5 display:decimal:format? [min|max|def] 102 – Fluke THE HART 1529 User Manual
Page 108: 6 display:decimal:format |min|max|def 102, Display:lamp? [min|max|def, Display:lamp ||min|max|def, Display:decimal:format? [min|max|def, Display:decimal:format |min|max|def

DISP:RES 3
The
parameter can be 0, 1, 2, 3, or 4. Specifying the MIN, MAX or DEF parameter
sets the display resolution to the minimum, maximum, or default value respec-
tively. Specifying the AUT parameter sets the display resolution to automatic.
8.4.6.3
DISPlay:LAMP? [MIN|MAX|DEF]
Returns the lamp brightness setting.
DISP:LAMP?
0
Specifying the MIN parameter returns a value of 0. Specifying the MAX pa-
rameter returns a value of 3. Specifying the DEF parameter returns a value of 0.
8.4.6.4
DISPlay:LAMP
Sets the lamp brightness setting.
DISP:LAMP ON
The lamp can be turned on or off by using either the
OFF), the
ters. Specifying the
MIN, MAX, or DEF sets the lamp to the minimum, maximum, or default value
respectively.
8.4.6.5
DISPlay:DECimal:FORMat? [MIN|MAX|DEF]
Returns the current decimal format setting.
DISP:DEC:FORM?
0
A returned value of 0 indicates a period (.) is used as the decimal separator. A
returned value of 1 indicates a comma (,) is used as the decimal separator.
Specifying the MIN parameter returns a value of 0. Specifying the MAX pa-
rameter returns a value of 1. Specifying the DEF parameter returns a value of 0.
8.4.6.6
DISPlay:DECimal:FORMat
Sets the decimal format setting.
DISP:DEC:FORM 1
The
Specifying the MIN, MAX or DEF parameter sets the decimal format to the
minimum, maximum, or default value respectively.
1529 Chub-E4 Thermometer Readout
User’s Guide
102