8.4.2.6

SENSe:RESistance:RANGe |MIN|MAX|DEF

Sets the resistance range setting.

SENS4:RES:RANG MIN

This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.

The SENSE suffix, , specifies the channel (1 to 4), or memory channel (5
to 14). The parameter must be a number between 0 and 1. Specifying
the MIN, MAX or DEF parameter sets the resistance range setting to the mini-
mum, maximum, or default value respectively.

This command may be password-protected. To access this command:

• If the SYST:PASS:CONV? command returns a 0 (probe password-protec-

tion is disabled), the password is not required.

• If the SYST:PASS:CONV? command returns a 1 (probe password-protec-

tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN command (see Section 9.4.11, Password Com-
mands).

8.4.2.7

SENSe:RESistance:WIRE? [MIN|MAX|DEF]

Returns the probe wire setting.

SENS1:RES:WIRE?

3

This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.

The SENSE suffix, , specifies the channel (1 to 4), or memory channel (5
to 14). Specifying the MIN parameter returns a value of 2. Specifying the
MAX parameter returns a value of 4. Specifying the DEF parameter returns a
value of 4.

8.4.2.8

SENSe:RESistance:WIRE |MIN|MAX|DEF

Sets the probe wire setting.

SENS2:RES:WIRE MIN

This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.

The SENSE suffix, , specifies the channel (1 to 4), or memory channel (5
to 14). The parameter must either be 2, 3, or 4. Specifying the MIN,
MAX or DEF parameter sets the wire setting to the minimum, maximum, or
default value respectively.

1529 Chub-E4 Thermometer Readout
User’s Guide

90