1 demand log 52, Figure 25 logging menu 52, Demand log 52 – Fluke THE HART 1529 User Manual
Page 58: Demand log, Figure 25, Logging menu, 159c, 142c, 634c, 015c
this menu are DEMAND LOG, AUTO LOG, LOG STATS, DATA LABELS,
and DEFAULT LABELS.
7.4.1
Demand Log
The DEMAND LOG submenu allows measurements to be logged on demand,
one reading at a time. Readings are stored to the instrument's memory and as-
signed a label for easy recall and organization of data. The instruments memory
can store up to 100 demand log readings.
Demand log values are stored exactly as they appear on the upper portion of the
display. The resolution, temperature units, and the channels displayed should
be set up prior to demand logging in order to ensure proper storage of data.
When this function is selected, the bottom portion of the display shows the de-
1529 Chub-E4 Thermometer Readout
User’s Guide
52
1
24.159C
3
24.142C
2
24.634C
4
24.015C
DATA RECORDING FUNCTIONS
DEMAND
LOG
AUTO
LOG
LOG
STATS
DATA
LABELS
DEFAULT
LABELS
EXIT
ENTER / MENU
UNIT
CONTRAST
1529
CHUB E-4 THERMOMETER READOUT
Figure 25 Logging Menu