4 system:password:conversion? [min|max|def] 117, 6 system:password:new 117, System:password:conversion? [min|max|def – Fluke THE HART 1529 User Manual
Page 123: System:password:conversion |min|max|def, System:password:new

SYST:PASS:CEN:STAT?
0
The response is either 0 (disabled) or 1 (enabled). A response of 0 indicates
password-protected settings cannot be accessed. A response of 1 indicates the
password was successfully entered and password-protected settings can be
accessed.
8.4.11.4
SYSTem:PASSword:CONVersion? [MIN|MAX|DEF]
Returns the current probe password-protected state.
SYST:PASS:CONV?
0
If the response to this command is 0, the indicated conversion settings are not
password-protected and can be accessed without having to enter the password.
If the response to this command is 1, the indicated conversion settings are pass-
word-protected and the password must be entered before they can be accessed.
Specifying the MIN parameter returns a value of 0. Specifying the MAX pa-
rameter returns a value of 1. Specifying the DEF parameter returns a value of 0.
8.4.11.5
SYSTem:PASSword:CONVersion
Sets the probe password-protection state.
SYST:PASS:CONV 0
The
protected (1 or ON) or not (0 or OFF). Setting this to 1 requires the password to
be entered before the indicated conversion settings can be accessed. Setting this
to 0 allows the indicated conversion settings to be accessed without having to
enter the password. This command is password protected. To access this com-
mand, the password must be successfully entered using the SYST:PASS:CEN
8.4.11.6
SYSTem:PASSword:NEW
Sets the current system password to a new password.
SYST:PASS:NEW 1212
The
is required to access the indicated settings. This command is password pro-
tected. To access this command, the password must be successfully entered us-
ing the SYST:PASS:CEN
Commands).
117
8 Digital Communications Interface
Commands