9 calculate:convert:snumber? 96, 10 calculate:convert:snumber 96, 11 calculate:convert:test? | 96 – Fluke THE HART 1529 User Manual
Page 102: Calculate:convert:snumber, Calculate:convert:snumber :convert:test? |

CALC:CONV:PRIN 2
“PRB”,CHAN2,”ser”,090105,”TYP”,RES,”RANGE”,0
If no parameters are specified, the ALL parameter is assumed and the parame-
ters for all channels are printed to the current port. Specifying the
rameter prints only the parameters for the specified channel (1 to 4) to the
current port. Optionally, the parameters may be printed to a specific port by
specifying the
8.4.4.9
CALCulate
Returns the probe serial number for the specified channel.
CALC2:CONV:SNUM?
A_336C
The CALCulate suffix,
nel (5 to 14). The returned serial number is in string format.
8.4.4.10
CALCulate
Sets the probe serial number for the specified channel.
CALC2:CONV:SNUM A_336C
The CALCulate suffix,
nel (5 to 14). The
can consist of up to eight characters that include any letters, numeric digits, and
the underscore ‘_’.
This command may be password-protected. To access this command:
• If the SYST:PASS:CONV? command returns a 0 (probe password-protec-
tion is disabled), the password is not required.
• If the SYST:PASS:CONV? command returns a 1 (probe password-protec-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN
mands).
8.4.4.11
CALCulate
Tests the temperature conversion for the specified channel.
CALC2:CONV:TEST? 100.0145
0.0100
The CALCulate suffix,
nel (5 to 14). For PRT/Thermistor channels, the
sistance value to convert to temperature. The
Ohms for PRTs and KOhms for Thermistors. For Thermocouple channels, the
1529 Chub-E4 Thermometer Readout
User’s Guide
96