• Sets the units to C.
• Sets the resolution to AUTO.
• Stops auto logging.
• Clears the statistical functions.

8.4.8.4

SYSTem:BOOT:VERSion?

Returns the boot ROM version.

SYST:BOOT:VERS?

0.50

8.4.8.5

SYSTem:CODE:VERSion?

Returns the firmware code version.

SYST:CODE:VERS?

1.11

8.4.8.6

SYSTem:ERRor?

Returns a system error message if any are present in the system error queue.

SYST:ERR?

0,“No error”

Each error condition produces only one error message at a time. The error
queue may contain up to ten messages. The messages are reported in the order
they occured. Reading a message from the queue removes the message so the
next message can be read. If more than ten errors occur before being read, the
tenth error in the queue is -350,“Queue overflow” and all but the first nine er-
rors are discarded. All errors cause the Error Bit (ERR, bit 2) for the Status
BYTE Register to be set. Refer to Troubleshooting section for the error
messages.

8.4.8.7

SYSTem:SNUMber?

Returns the instrument serial number.

SYST:SNUM?

A09002

If the serial number has not been set, a 0 is returned. The serial number is set at
the factory and cannot be changed!

8.4.8.8

SYSTem:VERSion?

Returns the SCPI version number.

SYST:VERS?

1529 Chub-E4 Thermometer Readout
User’s Guide