ITC Hemochron Signature Elite Whole Blood Microcoagulation System User Manual
Page 22

22
Specifying Continuous POCT-1A Communication
Note: Communication is initiated from the instrument only when the CLSI POCT-1A compliant 
communication protocol is used. If the ITC proprietary communication protocol is used, 
communication is initiated from a personal computer or network. 
1.
From the Supervisor menu, press 7 to display the Communications menu:
2.
Press 3 to select the desired communication option when the CLSI POCT-1A compliant 
communication protocol is used. (Display OFF to initiate communication only if requested, or 
display ON to initiate continuous communication whenever the instrument is started.) 
Note: Specifying the communication option acts as a toggle. If initiation of communication upon 
startup is already specified (Push POCT ON is displayed), this can be cancelled by pressing 3 again 
to display Push POCT OFF. 
Entering Network Communication Parameters
Some or all of the following parameters may need to be entered in order for the instrument to 
communicate over a network: 
•
IP Address
•
Gateway
•
NetMask
•
Remote IP
•
Remote Port
 Note: Consult with the network administrator for your facility to determine which parameters 
must be entered and what those settings should be. 
1.
From the Supervisor menu, press 7 to display the Communications menu:
2.
(If the CLSI POCT-1A compliant communication protocol is currently selected): Press 1 to toggle the 
communication protocol to the ITC proprietary communication protocol: 
Note: The CLSI POCT-1A compliant communication protocol must be OFF when running XPORT 
Setup. 
3.
Press 4 to start the XPORT Setup module. The message XPORT Setup Please Wait is displayed, 
followed by the first page of the XPORT Setup module: 
Note: If the message XPORT Setup NOT RESPONDING! is displayed, contact ITC Technical Support 
at (800) 631-5945 or (732) 548-5700, or by e-mail at [email protected]. 
