beautypg.com

RIGOL DS1000D User Manual

Page 66

background image

RIGOL

© 2008 RIGOL Technologies, Inc.

User’s Guide for DS1000E, DS1000D Series

2-36

Edge Trigger

Edge Trigger Mode; An edge trigger determines whether the oscilloscope finds
the trigger point on the rising or the falling edge of a signal. Select Edge trigger
Mode to trigger on Rising edge, falling edge or rising & falling edge.

Press Trigger MENU→Mode→Edge to display the edge trigger menu as follows.

Figure 2- 42 Table 2- 20 The Edge Trigger Menu



Menu

Settings

Comments

Source

CH1
CH2
EXT
AC Line
D15-D0

Select CH1 as trigger signal
Select CH2 as trigger signal
Select EXT TRIG as trigger signal
Select power line as trigger signal
Select a digital channel in D15-D0 as
trigger source (for DS1000D series)

Slope

Rising
Falling

Rising &

Falling

Trigger on rising edge
Trigger on falling edge
Trigger on both ring & falling edge

Sweep

Auto

Normal
Single

Acquire waveform even no trigger
occurred
Acquire waveform when trigger occurred.
When trigger occurs, acquire one
waveform then stop

Set up

To go to Set Up menu, see table 2-38