User’s Guide for DS1000E, DS1000D Series

1-9

 Align the slot in the probe connector with the button on the CH1 BNC.
 Push to connect, and twist to the right to lock the probe.
 Attach the probe tip and ground lead to the PROBE COMP connector.

Figure 1- 10

Attach the probe

Set the probe attenuation to 10X. To do this, push CH1→Probe→10X.

Figure 1- 11 Figure 1- 12
Set attenuation on the probe Set attenuation in the menu

Push the AUTO button. Within a few seconds, a square wave will display.


Push the CH1 button again to turn off Channel1.

Push the CH2 button to turn on channel2, repeat steps 2 and 3.


NOTE:
The signal output from Probe compensator should only be used for probe
compensation, not for calibration.

Probe compensator