Example 5: Making Cursor Measurements

There are 22 build-in automatic measurements. They can also be conducted using
cursors to make time and voltage measurements of a waveform quickly.

Measure the Peak Frequency of the First Sinc Waveform

To measure the ringing frequency at the rising edge of a signal, do these steps:

1. Press Cursor key to see the Cursor menu.

2. Press Mode to set Manual mode.

3. Press Type to select X.

4. Turn ( ) knob to place cursor A on the first peak of the wave.

5. Turn ( ) knob to place cursor B on the second peak of the wave.

Observe the delta in time and frequency displayed on the screen.

Figure 3- 4

Waveform display