To Compensate Probes

Perform this adjustment to match the characteristics of the probe and the channel
input. This should be performed whenever attaching a probe to any input channel for
the first time.

1. From CH1 menu, set the Probe attenuation to 10X (press CH1→Probe→10X).

Set the switch to 10X on the probe and connect it to CH1 of the oscilloscope.
When using the probe hook-tip, inserting the tip onto the probe firmly to ensure
a proper connection.
Attach the probe tip to the Probe compensator connector and the reference lead
to the ground pin, Select CH1, and then press AUTO.

2. Check the shape of the displayed waveform.

Over compensated Correctly Compensated

Under Compensated

Figure 1-9


Figure 1- 13

Probe Compensation


3. If necessary, use a non-metallic tool to adjust the trimmer capacitor on the probe

for the flattest square wave possible as displayed on the oscilloscope.

4. Repeat as necessary.

WARNING: To avoid electric shock while using the probe, be sure the

perfection of the insulated cable, and do not touch the metallic portions of

the probe head while it is connected with a voltage source.

!