Specifications, General – Precision Digital PD6730X User Manual
Page 8
PD6730X Vantageview Super Snooper Modbus Scanner
Instruction Manual
8
SPECIFICATIONS
Except where noted all specifications apply to operation at +25°C.
General
DECIMAL DISPLAY
(-2 Models)
Five Digits Top Display
(0 to 99999)
0.7" (17.8 mm) high, 7-segment,
automatic lead zero blanking.
Seven Characters
Bottom Display
0.4" (10.2 mm) high, 14-segment, automatic lead zero
blanking.
Symbols
Total, grand total, high alarm, low alarm, SafeTouch
button sleep mode/disable, password lock
FEET & INCHES
DISPLAY
(-L Models)
Feet & Inches Top
Display
0.60" (15.2 mm) high
0 to 399FT 1115/16 IN
7-segment, programmable 1/16 or 1/8 fraction display
Seven Characters
Bottom Display
0.4" (10.2 mm) high
14-segment, 7-digits
Tank Level Indicator
20-segments, F (Full) and E (Empty)
Alarm Indication
High and low alarm
Backlight White
DISPLAY
ASSIGNMENT
Top and Bottom Display*: Process Variables (PV); Alternating PV and Units, Tag
and PV, or Tag, PV, and Units; Pulse Input Rate, Total, or Grand Total with
Alternating Tag.
Bottom Display: All Top Display Options or Off
Units and tag independent for each PV, pulse input rate, total, and grand total.
* On feet and inches display models, top display used only for level Modbus
process variables or math channels.
BACKLIGHT
Backlight deactivated below temperatures ≈ -20°C
ALARM INDICATION Flashing display plus HI/LO (alarm) or SET indicators
SCAN AND
UPDATE RATE
Ambient > -20°C: Modbus PV scan rate programmable from 2 to 99 seconds per
PV. Tag and units programmable for 1 to 5 second alternation. Pulse input
variables update 1/second. Rate update is dependent on gate settings.
Ambient < -20°C: All Modbus scan, alternating units and tags, and pulse input
variables update/10 seconds minimum.
UNDERRANGE
Upper Display: Decimal display flashes -9999
Level display flashes to 399
FT
11
15
/
16
IN
Lower Display: Flashes -999999
OVERRANGE
Upper Display: Decimal display flashes 99999
Level display flashes to 399
FT
11
15
/
16
IN
Lower Display: Flashes 9999999
PROGRAMMING
METHODS
Four SafeTouch
®
through-window buttons when cover is installed. Four internal
pushbuttons when cover is removed.
PASSWORD MENU
OPTIONS
Three programmable password selections can be used for the following: restrict
modification of settings, prevent resetting the total or grand total without the
password, or permanently lock out the ability to change or reset the grand total or
any grand total related settings (making a non-resettable grand total).
Pass: Restricts modifications of programmed settings to require re-entering the
password to make changes.
Pass T: Restricts the reset of total to require re-entering the password. Disables
the manual mode reset contact.
Pass GT: Restricts the reset of grand total to Require re-entering the password.
May enable a non-resettable grand total and permanent lockout of grand total-
related settings with a specific password.