Calibrate o2 sub-menu, Setup sub-menu, Calibrate o – Emerson Process Management 3000 User Manual
Page 74: Sub-menu -4, Setup sub-menu -4, Table 5-3, Probe data sub-menu -4, World class 3000
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
5-4 General User Interface (GUI) Operation
Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5-8
CALIBRATE O
2
SUB-MENU
The CALIBRATE O
2
sub-menu (Table 5-4) is
used to enter the calibration mode. To access
the CALIBRATE O
2
sub-menu, press the CAL
key at any time. The increase and decrease
keys are used to scroll through the list.
The CALIBRATE O
2
sub-menu has three se-
lections available: Perform Calibration, View
Constants, and Calibration Status. Refer to
Table 5-4 for contents of the sub-menus.
Perform Calibration has two options depending
on how Auto Cal is selected in Probe Setup.
Refer to SETUP Setting in Table 5-4.
For information on performing a calibration, re-
fer to Section 4, Calibration.
5-9 SETUP
SUB-MENU
The SETUP sub-menu is used to enter all op-
erator set variables into the system. To access
the SETUP sub-menu press the SETUP key at
any time. To select the parameter to be
changed, move the cursor to the desired pa-
rameter using the arrow keys. Press ENTER to
select that parameter. To change the value for
that parameter, use the arrow keys to increase
or decrease the value. Press ENTER to save
changes.
There are six selections available on the
SETUP sub-menu: Calibration, O
2
Calculation,
O
2
Alarms, Efficiency Calc., Relay Outputs, and
Analog Outputs. Refer to Table 5-5 for the con-
tents of the SETUP sub-menu, or ESCAPE to
abort changes.
Table 5-3. PROBE DATA Sub-Menu
SUB-MENU
SELECTION
PARAMETER
DESCRIPTION
Process Data
O
2
__% O
2
O
2
value for the probe.
Efficiency
__%
Efficiency display.
Stack Temp
__DegC
Stack temperature.
Diagnostic Data
Temperature
Cell
__DegC
Cell temperature of the probe.
Stack
__DegC
Stack temperature.
Cold Junct
__DegC
Cold Junction temperature.
Voltages
Cell
__mV
Cell voltage of the probe.
Cell T/C
__mV
Cell thermocouple voltage of the probe.
Stk T/C
__mV
Stack thermocouple voltage.
Cold Jnt
__mV
Cold junction voltage.
Output Values
Analog
__% FS
Analog output voltage.
K1 State
OFF/ON
Status of relay 1.
K2 State
OFF/ON
Status of relay 2.