beautypg.com

Calibrate o2 sub-menu, Setup sub-menu, Calibrate o – Emerson Process Management 3000 User Manual

Page 74: Sub-menu -4, Setup sub-menu -4, Table 5-3, Probe data sub-menu -4, World class 3000

background image

Instruction Manual

IB-106-300NH Rev. 4.2
July 2002

5-4 General User Interface (GUI) Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

5-8

CALIBRATE O

2

SUB-MENU

The CALIBRATE O

2

sub-menu (Table 5-4) is

used to enter the calibration mode. To access
the CALIBRATE O

2

sub-menu, press the CAL

key at any time. The increase and decrease
keys are used to scroll through the list.

The CALIBRATE O

2

sub-menu has three se-

lections available: Perform Calibration, View
Constants, and Calibration Status. Refer to
Table 5-4 for contents of the sub-menus.

Perform Calibration has two options depending
on how Auto Cal is selected in Probe Setup.
Refer to SETUP Setting in Table 5-4.

For information on performing a calibration, re-
fer to Section 4, Calibration.

5-9 SETUP

SUB-MENU

The SETUP sub-menu is used to enter all op-
erator set variables into the system. To access
the SETUP sub-menu press the SETUP key at
any time. To select the parameter to be
changed, move the cursor to the desired pa-
rameter using the arrow keys. Press ENTER to
select that parameter. To change the value for
that parameter, use the arrow keys to increase
or decrease the value. Press ENTER to save
changes.

There are six selections available on the
SETUP sub-menu: Calibration, O

2

Calculation,

O

2

Alarms, Efficiency Calc., Relay Outputs, and

Analog Outputs. Refer to Table 5-5 for the con-
tents of the SETUP sub-menu, or ESCAPE to
abort changes.

Table 5-3. PROBE DATA Sub-Menu

SUB-MENU

SELECTION

PARAMETER

DESCRIPTION

Process Data

O

2

__% O

2

O

2

value for the probe.

Efficiency

__%

Efficiency display.

Stack Temp

__DegC

Stack temperature.

Diagnostic Data

Temperature

Cell

__DegC

Cell temperature of the probe.

Stack

__DegC

Stack temperature.

Cold Junct

__DegC

Cold Junction temperature.

Voltages

Cell

__mV

Cell voltage of the probe.

Cell T/C

__mV

Cell thermocouple voltage of the probe.

Stk T/C

__mV

Stack thermocouple voltage.

Cold Jnt

__mV

Cold junction voltage.

Output Values

Analog

__% FS

Analog output voltage.

K1 State

OFF/ON

Status of relay 1.

K2 State

OFF/ON

Status of relay 2.