beautypg.com

World class 3000, Instruction manual – Emerson Process Management 3000 User Manual

Page 67

background image

Instruction Manual

IB-106-300NH Rev. 4.2

July 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Calibration 4-7

World Class 3000

PROBE

(END VIEW)

IFT

CALIBRATION GAS

HPS

OPTIONAL

CHECK

VALVE

PROBE

SIGNAL CONNECTIONS

MPS-IFT

SIGNAL

CONNECTIONS

MPS

INSTRUMENT

AIR IN

CALIBRATION

GAS 1

(HIGH

)

O

2

CALIBRATION

GAS 2

(LOW O )

2

27270006

NOTE: THE MPS CAN BE USED WITH UP

TO FOUR PROBES. ONLY ONE PROBE

CAN BE CALIBRATED AT A TIME.

PROBE CALIBRATIONS MUST BE

SCHEDULED IN MULTIPLE PROBE

APPLICATIONS.

NOTE: SHOWN WITH HPS OPTION.

REFERENCE

AIR

Figure 4-4. Typical Automatic Calibration System

4