World class 3000, Instruction manual – Emerson Process Management 3000 User Manual
Page 154
Instruction Manual
Appendix E Rev. 4.5
June 1999
E-6 Appendices
Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
YES
YES
OXYGEN ANALYZER
SYSTEM IS EQUIPPED
WITH HPS 3000.
CHECK JUMPERS JM6 ON
MICROPROCESSOR BOARD
AND JM1 ON INTERCONNECT
BOARD ARE INSTALLED.
CHECK JUMPERS 9
OR 10 ON POWER
SUPPLY BOARD FOR COR-
RECT CONFIGURATIONS.
REFER TO PROBE
TROUBLESHOOTING,
APPENDIX A, SECTION 2.
CHECK THAT JUMPERS
JM6 ON MICROPROCESSOR
BOARD AND JM1 ON
INTERCONNECT BOARD
ARE REMOVED.
REFER TO APPENDIX B,
SECTION 2, FOR
JUMPER LOCATIONS.
DISCONNECT POWER
AND INSTALL JUMPERS
CORRECTLY. REFER TO
FIGURE 2-5 OF MAIN IB
SECTION.
NO
NO
SYMPTOM - MICROPROCESSOR BOARD LED IS STEADY ON
27270021
Figure E-4. IFT Troubleshooting Flowchart, #2
END OF FLOWCHART #2
SYMPTOM - GENERAL USER INTERFACE OR
LED DISPLAY PANEL DOES NOT FUNCTION
REPLACE GUI BOARD.
STILL NOT FUNCTIONING.
CALL FACTORY FOR
ASSISTANCE.
27270020
Figure E-5. IFT Troubleshooting Flowchart, #3